Calculate Vickers hardness from test force and indentation diagonals, with unit conversion, diagonal mismatch checks and a calculation summary.
Advanced options
How to use our Vickers Hardness Number Calculator
- Choose Force input mode so your Test force (selected unit) is entered in kgf or N correctly.
- Enter the applied load in Test force (selected unit). Use a positive value only.
- Choose Diagonal input mode. Pick Two diagonals if you measured both directions, or Mean diagonal only if you already averaged them.
- If using two diagonals, enter Diagonal 1 (mm) and Diagonal 2 (mm). If using mean mode, enter Mean diagonal (mm).
- Open Advanced options to change "Summary load unit", "Number display format", "Decimal places (digits)" or "Diagonal mismatch warning threshold (percent)".
- Click Calculate to see "Vickers hardness number (HV)", the mean diagonal, SI-equivalent hardness and "Calculation summary".
- Sanity-check the result: a smaller diagonal at the same force should give a higher Vickers hardness number (HV), and the Mean diagonal shown should match the value you expected.
- If you used two diagonals, review Diagonal mismatch and any Measurement warning. A large mismatch means you should re-check the indent image, focus, polish, or units before reporting.
Definitions
Force input mode: The unit used for the applied test force, either kgf or N.
Test force (selected unit): The load applied by the Vickers indenter during the test.
Diagonal input mode: Chooses whether you enter both measured diagonals or one already-averaged diagonal.
Mean diagonal: The average indentation diagonal length used in the hardness calculation. In two-diagonal mode, it is the arithmetic mean of Diagonal 1 and Diagonal 2 [4][3].
Vickers hardness number (HV): The classic Vickers hardness result. Higher HV means a harder surface under the same test method.
Diagonal mismatch: The percent difference between the two measured diagonals relative to their mean. Larger values can point to an uneven indent or measurement issue.
SI equivalent hardness: The same hardness result expressed as pressure, shown in MPa or GPa.
Calculation summary: Hardness and test force shown as separate quantities. It is not a complete test-report designation; conventional HV notation uses a kgf-based load designation and may also require dwell time.
Common mistakes and quick fixes
Mistake: Entering newtons while Force input mode is still set to kgf, or the reverse.
Fix: Match Force input mode to the unit used in Test force (selected unit) before you calculate.
Mistake: Typing diagonal values in um but leaving Diagonal 1 (mm) , Diagonal 2 (mm) , or Mean diagonal (mm) in millimeters.
Fix: Convert your microscope reading to mm first. For example, 500 um must be entered as 0.500 mm.
Mistake: Filling Mean diagonal (mm) while Diagonal input mode is set to two diagonals.
Fix: In two-diagonal mode, enter Diagonal 1 (mm) and Diagonal 2 (mm) . In mean mode, enter only Mean diagonal (mm) .
Mistake: Ignoring a large diagonal mismatch because a number was calculated. Fix: Read "Measurement warning" and recheck the indentation or measurement before using the result.
Mistake: Copying a calculation summary as a complete test designation. Fix: Use your applicable test standard and laboratory reporting requirements, including test force and dwell time; the summary alone is not a test report.
Mistake: Assuming SI equivalent hardness in MPa or GPa is a different test result from Vickers hardness number (HV) .
Fix: Treat the MPa and GPa values as conversions of the same hardness result, useful when you need SI-style pressure units.
Limitations & Key Assumptions / Boundary Conditions
- This calculator assumes a standard Vickers test using the usual 136 degree pyramid geometry and the classic HV equations.
- Results are only as good as the measured diagonal lengths. Poor focus, rough polish, edge effects, or an irregular indent can change the hardness value.
- In Mean diagonal only mode, the calculator cannot compute Diagonal mismatch because it does not know the two original measurements.
- The MPa and GPa outputs are SI-equivalent conversions of HV, not separate test methods.
- Very small diagonal values can produce extremely large HV results. When that happens, verify the microscope scale, decimal place, and force unit carefully.
- The calculation summary separates hardness from force. It does not verify test conditions or supply the complete notation and details required by a laboratory, customer or standard.
- If Diagonal mismatch warning threshold (percent) is set too low, you may see frequent warnings; if set too high, useful quality-control flags may be missed.
Methodology
Calculation steps
The calculator first determines which force unit you entered. If you enter force in newtons, it converts that value to kilogram-force using standard gravity before calculating HV.
d = (d1 + d2) / 2
F_kgf = F_N / 9.80665
HV = 1.8544 x F_kgf / d^2
Hardness_MPa = HV x 9.80665
Hardness_GPa = Hardness_MPa / 1000
Mismatch_pct = abs(d1 - d2) / d x 100
What each part means
d is the mean diagonal in mm. In two-diagonal mode, it is the average of the two measured indentation diagonals, which is the standard basis for the calculation [1][3]. F_kgf is the test force in kilogram-force. The constant 1.8544 is the standard geometry constant for the Vickers indenter.
Worked mini-example
Suppose Force input mode is kgf, Test force (selected unit) is 10, and both diagonals are 0.500 mm.
d = (0.500 + 0.500) / 2 = 0.500 mm
HV = 1.8544 x 10 / 0.500^2 = 74.176
Hardness_MPa = 74.176 x 9.80665 = 727.273 MPa
Hardness_GPa = 727.273 / 1000 = 0.727273 GPa
The summary is about 74.18 HV with test force 10 kgf, or the equivalent 98.07 N. Changing "Summary load unit" changes only the separate force label; it does not put a Newton value into the conventional HV load designation.
Warnings and interpretation
If you enter two diagonals, the calculator also computes diagonal mismatch. A larger mismatch does not automatically make the HV wrong, but it is a useful quality-control flag because the two diagonals are often slightly distorted in practice [2]. Higher HV means a harder material surface for that test condition.
Assumptions used
This tool assumes positive force and positive diagonal values, standard gravity of 9.80665 N per kgf, and consistent millimeter input for diagonal lengths. It does not adjust for dwell time, surface preparation quality, indenter condition, or lab-specific reporting rules, all of which can affect real test results.